Owner’s ManualMedical-Grade Mobile Power Retrofit KitModel #: HCRK-INTKit Includes: Power Supply Module, Battery Module, Remote User Interface (RUI),P
10FeaturesRemote User Interface (RUI):“Power” Button: This button turns the Power Supply Module’s outlets(and any connected equipment) on and off.To t
11Features continuedPower Supply Module continued:USB Port: This port connects the Power Supply Module to any cart-mounted computer (laptop or desktop
12TroubleshootingProblemNo AC output power available at outlets.Battery Module not recharging,even with AC utility powerpresent.Low battery alarm soun
13Storage & ServiceSafety & Regulatory StatementsStorageFor short term storage (when charging the Battery Module between shifts), see “Operati
14Warranty Limited WarrantyTripp Lite warrants its Power Supply Module to be free from defects in materials and workmanship for a period of 2 years fr
15200702102 93-2657 HCRKINT Retrofit Kit OM.qxd 4/27/2007 11:33 AM Page 15
161111 W. 35th Street, Chicago, IL 60609 USA+1 773 869 1234 • www.tripplite.com200702102 93-2657200702102 93-2657 HCRKINT Retrofit Kit OM.qxd 4/27/2
2Important Safety InstructionsTripp Lite’s Medical-Grade, Mobile Power Retrofit Kit includes two modules: a Power Supply Module anda Battery Module. W
3Battery Module Warnings• Do not allow objects to contact the Battery Module’s two DC input terminals. Do not short or bridge these terminals together
Power Supply Module & Battery Module MountingAttach the Power Supply Module to the mounting bracket using the included hardware.* Make sure the Po
Remote User Interface (RUI) MountingMount the RUI to the top edge of the mounting bracket with included screws.Or, if desired, use the included double
6Remote User Interface (RUI) MountingMount the Remote User Interface in a variety of positions on any flat surface using included double-sided tape.CA
7Connection / Start-UpConnect the Remote User Interface (RUI) to the Power Supply ModuleThe Power Supply Module will not operate without the RUI conne
8Connection / Start-Up continuedTurn the Power Supply Module’s Outlet ONPress and hold the “Power” button* for one second. The alarm will beep once br
Stationary Operation (Plugged In / Battery Charging)Plug the Power Supply Module into a live AC wall outlet.* The Power Supply Module will deliver ACp